Guidance and Manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an
environment.
Immunity
IEC 60601
test
test level
Electrostatic
6kV contact
discharge
8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
2kVfor power
transient/
supply lines
burst
1kVfor input/
IEC 61000-4-4
output lines
Surge
1kV line(s) to
IEC 61000-4-5
line(s)
2kV line(s) to
earth
interruptions
<5 % UT
and
(>95 % dip in UT)
voltage varia-
for 0,5 cycle
tions on
40 % UT
power supply
(60 % dip in UT)
input lines
for 5 cycles
IEC 61000-
4-11
70 % UT (30 %
dip in UT) for 25
cycles
<5 % UT (>95
% dip in UT) for
5 sec
Power
3 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the AC mains voltage prior to application of the test level
EN
Compliance level
6kV contact
8 kV air
2kVfor power
supply lines
1kV for input/
output lines
1 kV line(s) to
line(s)
2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT
(30 % dip in UT) for
25 cycles
<5 % UT (>95 % dip
in UT)
for 5 sec
3 A/m
Electromagnetic environ-
ment – guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the rela-
tive humidity should be at
least 30 %
Mains power quality should
be that of a typical commer-
cial or hospital environment
Mains power quality should
be that of a typical commer-
cial or hospital environment
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the device
requires continued operation
during power mains interrup-
tions, it is recommended that
the device be powered from
an uninterruptible power
supply or a battery
Power frequency magne-
tic fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment