Continence Treatment Programs
Programs
PAIN RELIEF
PR
STRESS 1
STR1
STRESS 2
STR2
URGE
URGE
FREQUENCY/URGE 10
FURG
PELVIC FLOOR
WORK OUT AND
LACK OF
SENSITIVITY
(34 MIN)
PFW
CUSTOMISED
MEM 1
CUSTOMISED
MEM 2
CUSTOMISED
MEM 3
The parameters in the table are suggested parameters only and should be used as that.
Please refer to the ACA Notes on Good Practice or The Chartered Society of
Physiotherapy Guidelines, to view the most up to date neuromuscular electrical
stimulation parameters.
As a clinician, you may prefer to use your own parameters to treat patients and it is
then ideal for you to use the customisable programs MEM1-3.
Hz
PW
Rate
μS
2
175
30
200
35
210
10
210
250
6 Sequential phases:
Phase 1: 2Hz for 4 min,
Phase 2: 10Hz for 10 min,
Phase 3: 15Hz for 5 min,
Phase 4: 20Hz for 5 min,
Phase 5: 30Hz for 5 min,
Phase 6: 10Hz for 5 min
All phases are 5 seconds work, 5 seconds rest at 220 µS,
0.8 seconds ramp.
2 - 100 50 - 450 0.3 - 9.9
2 - 100 50 - 450 0.3 - 9.9
2 - 100 50 - 450 0.3 - 9.9
PERICALM™ Instruction Manual - GB
Ramp up
Work
time in
time in
seconds
seconds
1
Cont
1
5
1
5
1
5
1
5
2 - 99
2 - 99
2 - 99
11
Rest
Overall
time in
time in
seconds
seconds
Cont
25 min
5
25 min
10
25 min
6
25 min
10
25 min
2 - 99
5 - 60
2 - 99
5 - 60
2 - 99
5 - 60