Connect the short silicon tube 1, with antibacterial filter 2, to the suction connector 3. The other tube,
with one end connected to the filter must be connected with the other end to the jar's connector 4
where has been fixed the red float (security float).
•
When the 90% of the volume of the jar is reached there is the activation of the security float (the float
close the aspiration connector on the jar) to avoid liquid penetration inside the device.
The device must be used on a plan of horizontal operation.
•
Connect the long silicon tube 5 to the other jar's lid connector 6
•
Connect the other end of the long silicon tube 5 to the probe plastic connector 7
•
Connect the power cord to the device than connect the plug to the electrical mains supply.
•
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and
an rapid reaching of the functionality vacuum) then rescrew the lid on the jar correctly.
•
Push switch 8 on position I to start suction
•
Once finished push switch on O position and unplug
•
To extract the accessories and start with cleaning it.
The dual use version is available on request.
Continuous use can be enabled by activating the illuminated switch.
Alternatively, the pedal control can be used for aspirating on command, while leaving the operator's hands
free. If you're using the pedal version, insert the plug of the pedal power cord into the appropriate socket
located next to the equipment, push the second switch on the position II, so start and stop the suction
operation by pressing relasind the pedal.