2. Establish the average for the lower section by carrying out
5 measurements using a torque meter.
5 x
5 x
90
Anzeige wechseln
Rekalibrierung_3
SI
Nm
ft·lbf
Nm
3. Pressing the display for over 8 seconds will change the electronic
system to Teach Mode for setting the lower value. You can see
that Teach Mode has been reached when the display alternates
between «CAL» (calibration low) and the display value. If you do not
touch the switch, the device will remain in this mode for 15 seconds
before the electronic system reverts to normal.
Please note: The display will change from «SI» to «IMP» after
3 seconds. To change to Teach Mode, continue to hold down the
display switch.
1.40
1.03
3 sec.
3 Sek.
SI
Nm
ft·lbf
Nm
IMP
ft·lbf
cal
5 sec.
8 Sek.
IMP
SI
ft·lbf
Nm
ft·lbf
91