sodium hypochlorite), specific for disinfection and is available in all pharmacies.
Directions:
-
Fill a container of suitable size to hold all the parts to be disinfected (we recommend the
use of the suitable Clenny
tion of drinking water and disinfectant, according to the proportions indicated on the pack-
aging of the disinfectant.
-
Completely immerse each individual component in the solution, taking care to avoid the
formation of air bubbles in contact with the components. Leave the components immersed
for the period of time indicated on the packaging of the disinfectant, and associated with
the concentration chosen to prepare the solution.
-
Remove the components now disinfected and rinse thoroughly with lukewarm drinking
water.
-
Dispose of the solution as recommended by the disinfectant manufacturer.
Method B: boil in water for 10 minutes; it is advisable to use demineralized water or distilled
water in order to avoid chalky deposits.
Method C: disinfect the accessories with an electrical hot steam steriliser for baby-bottles. Per-
form the process thoroughly following the instructions of the steriliser. To ensure the disinfec-
tion is effective, choose a steriliser with an operating cycle of at least 6 minutes.
Method D: disinfect the accessories in the microwave oven using the suitable Clenny
Box container, available as an accessory, following the instructions contained therein.
After disinfecting the accessories, shake them well without scrubbing them with clean cloths
and place them on a paper towel. Alternatively, dry them with a jet of hot air (for example a
hair dryer).
3.4 - Sterilization:
Sterilization intended for clinical or hospital use is the responsibility of the same structure that it
performs and must be performed according to the instructions in this manual.
The accessories that can be sterilized are (2-3-4)
The sterilization procedure described in this paragraph is effective only if all the prescriptions
herein are met, all the components involved are previously sanitized and the procedure itself is
validated according to ISO 17665-1 rule.
Unit: Steam sterilizer with fractionated vacuum and overpressure according to EN 13060 rule.
Procedure: Package each component to be treated in sterile barrier system or packaging complying
with EN 11607 rule. Place the packed components in the steam sterilizer. Run the sterilization cycle
according to the instructions manual of the device, selecting a temperature of 134° C and a duration
of 10 minutes.
Storage: Store sterilized components as per instructions of the sterile barrier package or system
chosen.
ADDITIONAL INFORMATION FOR CLEANING the nebulisation chamber (with the exception of the
compressor chamber)
To prevent deposits from obstructing the nebuliser and the nebulisation chamber and thus affecting
the operation of the nasal irrigator, we recommend cleaning and sanitising after each use by follow-
ing the operating instructions manual that is supplied with the nasal irrigator. After disinfecting the
accessories, shake them well without scrubbing them with clean cloths and place them on a paper
towel. Alternatively, dry them with a jet of hot air (for example a hair dryer).
At the end of each use store the device with its accessories in a dry place away from dust.
WARNING: Should your nasal irrigator not atomize properly after cleaning and sanitizing is per-
formed, do not use sharp tools (i.e. screwdrivers, scissors ends, needles, hairpins, etc.) to try to
clean or unclog the atomizer or atomization chamber holes as this could damage these parts and
impair the correct operation of the unit. Instead, perform additional cleaning and sanitizing cycles
as described in the instruction manual and above recommended, until the nasal irrigator starts to
operate properly again.