Guidance and manufacturer's declaration - electromagnetic emissions
The SIL 45 is intended for use in the electromagnetic environment specified below.
The customer or the user of the SIL 45 should assure that it is used in such an environment.
Emissions
RF emissions
CISPR 11(EN 55011)
RF emissions
CISPR 11 (EN 55011)
Harmonic emissions
IEC 61000-3-2
(EN 61000-3-2)
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
(EN 61000-3-3)
Guidance and manufacturer's declaration - electromagnetic immunity
The SIL 45 is intended for use in the electromagnetic environment specified below. The customer or the user of the SIL 45
should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
(EN 61000-4-2)
Electrical fast transient/
burst IEC 61000-4-4
(EN 61000-4-4)
Surge
IEC 61000-4-5
(EN 61000-4-5)
Voltage dips, short interrup-
tions and voltage variations
on power supply
input lines
IEC 61000-4-11
(IEC 61000-4-11)
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
(IEC 61000-4-8)
NOTE U
is the a.c. mains voltage prior to application of the test level
T
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Compliance
Group 1
The SIL 45 uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
Class B
The SIL 45 is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic pur-
Class A
poses.
Complies
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV line(s) and neutral
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5s
3 A/m
Electromagnetic environment - guidance
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV line(s) and neutral
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5s
Not applicable
31
Electromagnetic environment
--guidance
Floors should be wood, con-
crete or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If a dips
or an interruption of mains po-
wer occurs, the current of the
SIL 45 may be dropped off from
normal level, it may be necessa-
ry to use uninterruptible power
supply or a battery.
Not applicable