Usage
Pwr
P1
P2
P3
P4
Control panel
1. Setting date and time
Press
to activate the date and time setting
Press
to start the date setting
Press
to switch between year/month/day
Press
to adjust year/month/day
Press
again to save the date setting and start the time setting
Press
to switch between hour/minutes/seconds
Press
to adjust hour/minutes/seconds
Press
again to save the time setting and go back to main menu
2. Manual dosing
Press
to go to the manual dosing setting
Press
to select the pump
Press
to start manual dosing
Press
to go back to main menu
3. Calibration
Before programming, it is essential that you calibrate the Dosing pump for
each pump head you intend to utilise.
To do so, you will need a 100ml graduated cylinder and RO/DI or distilled
water.
In the main menu press
Press
to select the pump
Press
to go to the calibration setting for the chosen pump
Ensure that the input tube of the pump that you wish to calibrate is
immersed in water. Position the graduated measuring cylinder so to catch
all water from the output tube of pump being calibrated.
Press
to start. Once 100 mL has been dispensed, press the
stop the pump.
Press
to save the data and go back to the calibration menu. Repeat the
previous steps for every pump head.
Note: before you calibrate ensure that there is no air in the hose by manual
dosing!
Press
to go back to main menu
to go to the calibration setting
again to
3
GB
Set Date & Time <
Set Up Program
Set uP New Date
2019-05-01
Set uP New Time
12:24:09
Manu Mode
1
2
3
4
Manu Mode
1
2
3
4
Calibration Pump
1
2
3
4
P1 Calibration
100ml= 200