MAINTENANCE

Use the information in this section to properly maintain the operation of your
AC/DC Current Probe.
BATTERY AND AC POWER ADAPTER
A CP6000 Series current probe uses a single 9 V battery. As the probe is a high
power product, please use the specified alkaline battery.
As the battery installed is drained, a significant gain error may occur. The green
ON Indicator LED will continue to light until a low battery voltage of 6.5 V is
reached. If probe gain errors are detected, replace the battery.
The included AC Power Adapter can be used to avoid gain error due to a low
battery. Use the AC Power Adapter when there is AC power available.
The CP6000 Series current probes have in their design a priority external power
circuit, therefore it is safe to use the AC Power Adapter while the battery is
installed. However, when using the AC Power Adapter for an extended period,
removal of the battery is recommended.

BATTERY INSTALLATION

(1) Remove the probe from the circuit.
(2) Open the battery compartment by loosening and removing the three screws
on the back of the probe. Lift off the back cover.
(3) Observing polarity, attach the new alkaline battery to the connector buttons
and place the battery into the compartment, as illustrated in Figure 5.
(4) Replace the back cover and lightly tighten the three cover screws.
CP6000MAN v01