berlinger Q-tag 2 plus Mode D'emploi page 15

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17
Q-tag
®
plus
The next push to the START button shows another temperature display, identifiable by the «°» sign. It is
accompanied by the flashing «max» sign for a maximum temperature recorded during the whole period of
a high alarm or «min» sign for a minimum during a low alarm respectively.
Up to three alarms are recorded per segment and can easily be read as described above. After showing the
last recorded alarm of any segment, the display automatically jumps to the maximum of the following seg-
ment with the next push to START.
®
How can Q-tag
Q-tag
®
plus has a «Test-Mode» which can be used prior to starting and after stopping the device but not during
the active measurement. This mode is activated by firstly pressing and holding the STOP button, then simul-
taneously pressing START and releasing both. The display shows for 10 times alternately the current ambient
temperature and all display segments. Thus, many important functions like time and temperature measure-
ment, overall and display functions can be tested easily.
Important: Too high ambient temperatures may be displayed in Test-Mode due to heating by hands.
Testmode: Displaytest
The Q-tag
®
plus device automatically falls back to its prior operating mode after the tenth test cycle.
plus be tested? (Test-Mode)
History-Mode: extreme temperature
recorded during alarm event
10 x
Current ambient temperature display
15

Publicité

Table des Matières
loading

Produits Connexes pour berlinger Q-tag 2 plus

Ce manuel est également adapté pour:

Q-tag 2r plus

Table des Matières