MEMORY CLEARANCE

After a measurement is finished or when the device
stands by, hold down button "M1" or "M2" for at
least 5 seconds, the display will show "CLR", which
means the stored reading for "M1" or "M2" is remo-
ved as Fig. 10.

ERROR AND LOW BATTERY INFORMATION

INDICATION POSSIBLE REASON CORRECTION METHODS
C Err
• The cuff is put on in-
• Movement of
• The cuff is not infla-
• Arrhythmia
The batteries are weak
correctly or the tube
plug is inserted too
loosely
arm/hand or talking
during measurement.
ted to necessary
pressure
• Make sure that cuff is put
on correctly and the tube
plug is inserted tightly
and repeat the measure-
ment
• Repeat the measurement
the following completely
recommendations of ma-
nual
• Repeat the measurement
with pumping cuff on 30-
40 mmHg above expected
systolic pressure
• Consult your personal
physician
Replace all 4 batteries with
new ones