7. Post-use Maintenance
After each patient maintain the product as follows.
WARNING
• Immediately after a treatment (within 1 hour), perform maintenance and then store the handpiece.
Failure to properly maintain the handpiece may cause infection, product failue, overheating leading
burn injuries.
CAUTION
• Follow local rules, regulations, and guidelines regarding the reprocessing of devices.
• Perform the following cleaning and disinfecting quickly after treatment (within 1 hour) to remove any residue.
• Steam sterilization is recommended for the product. The validity of other sterilization methods (such as plasma sterilization
or EOG sterilization) is not confirmed.