"Hold" Display

Purpose
The last stable value remains displayed for 5 seconds after removing the
sample from the pan.
Example: Determine weight of oversized sample
Step
1. Select application program
2. Select Display Hold ("
3. Confirm setting
Symbol " " on the display:
application is active
4. If necessary: zero the balance/scale
5. Place sample on balance/scale
6. Start application program
Symbol "
" flashes on the display:
the weight value is locked
7. Unload the balance/scale:
the weight value remains displayed
for a further 5 seconds; or
8. Zero the balance/scale
9. End the Display Hold application
10. Reactivate Display Hold (if no other
application program has been selected)
11. Repeat procedure starting from Step 5.
12
HLdon
")
Key (or instruction)
(ZERO) ≥ 2 sec
(F) repeatedly
(ENTER)
(ZERO)
(F)
(ZERO)
(ENTER) ≥ 2 sec
(F)
Display