When illuminating heat-sensitive or flammable light-absorbing objects, pay
particular attention to selecting the light guide spacing and intensity so that
no thermal damage occurs to the object. Since light-absorbing materials
have the physical property of converting incident light into heat, damage can
occur to heat-sensitive or flammable light-absorbing materials.
All light guide outputs must always be at a safe distance – at least 10 cm –
from heat-sensitive or flammable light-absorbing materials when the light
source is switched on (avoid possible fire hazard).
Reduce the brightness and duration of illumination of biological tissue to
avoid unnecessary exposure of biological tissue to visible light illumination.
Keep all ventilation openings free. In case of insufficient cooling, an integra-
ted thermal switch will regulate the light intensity of the light source.
Safe disconnection from the power supply is only achieved by pulling out
the plug.
Do not open or disassemble the device. Refrain from making technical modi-
fications to the device.
Allow the filter slider and insert filter to cool down before removing the insert
filter. The filter slider must always be in a detent position when operating the
light source (see "Filter slide").
Allow the light guide to cool down before replacing it. We recommend that
you switch off the light source for about five minutes before replacing the
light guide.