Heraeus-Vôtsch
Operating
Instruction
Datasheet
for
Test-Chamber
ht7010/20
(ht7104)
Page
Rev.
.. os
un
Get
values:
compr.air
free
outl
free
out2
Direction
No.
ASCII
Meanin
Master
to
slave
$
Starting-sign
aa
Adress
(2 digits)
I
Command
«CR»
End
Slave
to
master
L
XXXX.X
Temper
-
Nominalvalue
2
XXXX.X
Temper
- Actualvalue
3
XXXX.X
fan
t.spa
-
Nominalvalue
4
XXXX.X
fan
t.spa
-
Actualvalue
5
XXXX.X
Pt100
1
-
Nominalvalue
6
XXXX.X
Pt100
1
- Actualvalue
7
XXXX.X
Pt100
2
-
Nominalvalue
8
XXXX.X
Pt100
2
-
Actualvalue
9
| XXXX.X
Pt100
3
-
Nominalvalue
10
XXXX.X
Pt100
3
- Actualvalue
11
XXXX.X
Pt100
4
-
Nominalvalue
12
XXXX.X
Pt100
4
~ Actualvalue
0
0
unused
1
b
start
2
b
Error
3
b
Temper
4
b
add.dehum
5
b
CO2
cool.
6
7
8
9
free
out3
b
b
b
b
10
b
free
out4
11
b
free
in
1
12
b
free
in
2
13
b
free
in
3
14
b
free
in
4
15
0
unused
<CR>
End
Example
(Master
to
slave):
SOOI<CR>
Example
(Slave
to
master):
0020.0
0020.0
0020.0
0020.0
0020.0
0020.0
0020.0
0020.0
0020.0
0020.0
0
20.0
0020.0
0111111111111110<CR>