Télécharger Imprimer la page

Tintometer Lovibond Water Testing MD100 Mode D'emploi page 36

Masquer les pouces Voir aussi pour Lovibond Water Testing MD100:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 48
Methods
GB
CL 10
Tablet
Powder
0.0.0
Tablet
Powder
Zero
Test
CL 10
Tablet
Powder
RESULT
Tablet
Powder
Zero
!
Test
CL 10
Tablet
Powder
RESULT
Tablet
Powder
36
Chlorine HR with DPD Tablet
0.1 – 10 mg/l
a) free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Remove the vial from the sample chamber and empty it, leaving a
few drops remaining in the vial.
Add one DPD No. 1 HR tablet straight from the foil to the water
sample and crush the tablet using a clean stirring rod.
Add the water sample to the 10 ml mark.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.
b) total Chlorine
Add one DPD No. 3 HR tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 2 minutes.
(Countdown can be activated, see page 31)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.
c) combined Chlorine
combined Chlorine = total Chlorine – free Chlorine
Tolerances:
0 – 2 mg/l: ± 0.1 mg/l
> 2 – 4 mg/l: ± 0.3 mg/l
> 4 – 8 mg/l: ± 0.4 mg/l
> 8 – 10 mg/l: ± 0.5 mg/l
marks
marks
MD100_11 04/2017

Publicité

loading