1-3 electrical Specifications
Accuracy is ±(% reading + number of digits) at
23°C ± 5°C < 80%RH.
Current
Range
Resolution
6mA *
0.001mA
60mA
0.01mA
600mA
0.1mA
6A
0.001A
60A
0.01A
100A
0.1A
* There is a little rolling within ±3 digits.
frequency Response (61~ 400Hz):
2%+6D(0~30A), 30%+6D(30~100A)
Influence of external magnetic field of
adjacent wires:
0.04% typical value.
Low Pass filter Mode
Range
Resolution
6mA *
0.001mA
60mA
0.01mA
600mA
0.1mA
6A
0.001A
60A
0.01A
100A
0.1A
* There is a little rolling within ±3 digits.
attenuation characteristic: approx. –24dB/oct
Cut-off frequency: 100Hz
ILCM 06 English
accuracy (50 ~ 60Hz)
0 ~ 50A: ±(1% + 3D))
50 ~ 60A: ±(5% + 6D)
60 ~ 100A: ±(8% + 6D)
accuracy (50 ~ 60Hz)
0 ~ 50A: ±(2% + 6D))
50 ~ 60A: ±(6% + 8D)
60 ~ 100A: ±(9% + 8D)