Télécharger Imprimer la page

Widex Dex COM-DEX Mode D'emploi page 83

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 23
All manuals and user guides at all-guides.com
Emissions test
Voltage fluctuations/
flicker emissions
IEC
61000-3-3
*) Battery powered equipment
Electromagnetic immunity
The COM-DEX Remote Mic is intended for use in the electromagnetic envi-
ronment specified below. The customer or the user of a COM-DEX Remote
Mic should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test level
Electro-static dis-
± 6 kV contact
charge (ESD)
± 8 kV air
IEC
61000-4-2
Compliance
Not
applicable *)
Compliance level Electromagnetic
± 6 kV contact
± 8 kV air
83
Electromagnetic envi-
ronment
- guidance
that supplies buildings
used for domestic pur-
poses.
environment
– guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are cov-
ered with syn-
thetic material,
the relative hu-
midity should be
at least 30 %.

Publicité

loading

Ce manuel est également adapté pour:

Com-dex remote mic