Troubleshooting and further Notes
 OH or OL is displayed when the temperature being measured is
outside the range of the instrument (OL: colder than -40 °C resp.
OH: hotter than +580 °C). Only select targets which are within the
specifications. If a malfunction still exists, please contact our cus-
tomer support.
 Observe the measuring distance/radius ratio in case the measur-
ing results are not comprehensible. Hold the thermometer closer
to the object to be measured.
 A big difference between the emissivity of the object to be
measured and the preset emissivity of the thermometer (0.95)
can lead to imprecise measurements.
 The thermometer should be allowed several minutes to stabilize
to the working temperature when it has been exposed to rapid
changes in the ambient temperature.
 Do not expose the thermometer to strong magnetic fields or static
charges. This will affect the performance of the thermometer.
 Make sure that there are no objects such as glasses, plastics,
vapour, dust, etc are in the space between the thermometer and
the object to be measured.
Cleaning and Storing
 Only use a soft, lit free lightly moistened cloth to clean the
housing of unit. Do not use aggressive chemicals or abrasives
and do not use liquids.
 Do not use force on the LC display when cleaning it.
 Carefully clean the infrared sensor with a soft brush or with clean
compressed air.
 If you do not plan to use the unit for some time, please remove
the battery.
 Store the unit indoors out of reach of children in a place that is
protected from moisture, dust, dirt and extreme temperatures.