Devices not to be used in presence of a mixture of inflammable anaesthetic with air or oxygen or nitrous
oxide.
CAUTION, in all cases:
WARNING: Always disconnect the electrical cord from the mains supply before opening the body of the
compressor (electrical shock hazard).
WARNING: Caution, only a professional repair shop is allowed to repair the device.
WARNING: Modification forbidden.
WARNING: Only use the accessories supplied and mentioned in the instructions for use.
WARNING: Do not undertake service or maintenance operations while the device is in use.
8. Failure remedies
Problem
No operating noise can be
heard.
No aerosol comes out of
the nebulizing kit.
9. List of accessories and spare parts
Accessories:
Extraneb+ nebulizing kit with adult mask and mouth piece, ref. NL9MXBR
Extraneb+ nebulizing kit with paediatric mask and mouth piece, ref. NL9MXBP
Extraneb+ nebulizing kit with adult mask, ref. NL9MXR
Extraneb+ nebulizing kit with paediatric mask, ref. NL9MXP
Extraneb+ nebulizing kit with mouth piece, ref. NL9MXB
Spare parts:
Mouth piece, ref. C27D
Adult mask, ref. R
Paediatric mask, ref. RP
Infant mask, ref. RBB
Tube 6x9, length 1 m, ref. C20P
Fuses: T 0,5A H 250V
Electric cord: OMEGA PZ1190020; length: 2 m; 2,5/10A – 250V; 0,75mm²
10. Warranty / recycling
Warranty: The breast pump is warranted 3 years, parts and labour, against any manufacturing defect, except case
of misuse. For after sales service under warranty, please contact your distributor. A technical description is
available on request.
Recycling
The device must be disposed of in accordance with local regulation on the disposal of electrical
appliances. So you contribute to the human health and environmental protection.
Product on the market since August the 13
Check that the electric cord is connected to the mains and to the device.
Check that the switch is in position on.
Check that main power supply is in accordance with the unit (Volt Hz).
Check that electrical current arrives to the socket trying with another device.
Check that the tank contains enough liquid depending on the jet nebulizing kit: the
minimum of medicine is indicated on the instructions for use of each nebulizing kit.
Check that the tube is well connected to the compressed air inlet of the nebulizing
kit and to the air outlet of the compressor.
Check that the tube is not pinched or blocked.
Check that the jet nozzle is completely inserted.
Jet nozzle must be cleared (clean it with water).
In case of use of medicine glass bulbs, check that glass fragments did not fall into
the nebulizer.
Solution
th
2005.
16