VIII.
TABLEAU DE CONFORMITE CEM
The NOCOSPRAY 2 is intended for use in the electromagnetic environment specified below.
The customer or the user of the NOCOSPRAY 2 should assure that it is used in such an environment.
Emission test
Emissions RF
Emissions RF
Harmonics emissions
Voltage/Flicker test
The NOCOSPRAY 2 is intended for use in the electromagnetic environment specified below. The customer or the user of the NOCOSPRAY 2
should assure that it is used in such an environment.
Immunity test
Electrostatic discharges
IEC 61000-4-2
Electrical Fast Transients
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
Magnetic field at power
frequency
IEC 61000-4-8
EMC Compliance according to IEC 60601-1-2 (2007)
Standard
Compliance
CISPR 11
Group 1
CISPR 11
Class B
IEC 61000-3-2
N/A
IEC 61000-3-3
N/A
IEC 60601
Compliance level
test level
6 kV contact
6 kV contact
8 kV air
8 kV air
2 kV for
2 kV for power
power
supply lines
supply lines
1 kV for
input/output
lines
1 kV line(s) to
1 kV line(s) to
line(s)
line(s)
2 kV line(s) to earth
2 kV line(s) to earth
<5 % U
<5 % U
T
T
(>95 % dip in
(>95 % dip in U
T
)
for 0,5 cycle
U
T
for 0,5 cycle
40 % U
T
40 % U
(60 % dip in U
)
T
T
(60 % dip in U
)
for 5 cycles
T
for 5 cycles
70 % U
T
70 % U
(30 % dip in U
)
T
T
(30 % dip in U
)
for 25 cycles
T
for 25 cycles
<5 % U
T
<5 % U
(>95 % dip in U
T
T
(>95 % dip in
for 5 s
)
U
T
for 5 s
3 A/m
3 A/m
Electromagnetic environment
The NOCOSPRAY 2 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The NOCOSPRAY 2 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low
voltage power supply network that supplies buildings used for domestic
purposes.
Electromagnetic environment
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
)
environment. If the user of the system requires continued operation
during power mains interruptions, it is recommended that the system be
powered from an uninterruptible power supply or a battery.
is the a.c. mains voltage prior to application of the test level.
U
T
)
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.