TENS PLUS - PREMIER STIM PLUS – Digital TENS/EMS
1. B Mode(Burst) Load: 500 ohm Pulse Rate: 150Hz Pulse Width: 300μs
Scope A :
VERT:10.0V/DIV
HORIZ:2mS
OUTPUT:57.1875Vpk
-pk
Pulse Rate:100Hz
2. N MODE(Normal): Load: 500 ohm Pulse Rate: 150Hz Pulse Width: 300μs
Scope A :
VERT:10.0V/DIV
HORIZ:2mS
OUTPUT:59.1825V pk-
pk
Pulse Rate:148.8Hz
3. M MODE(-50% Pulse Width & Rate Modulation): Load:500 ohm Pulse Rate:150Hz Pulse Width:300μs
Scope A:
VERT:10.0V/DIV
HORIZ:100μs
OUTPUT:54.9975Vpk
-pk
Pulse width:300μs
4. S1 MODE(-40% Pulse Width Modulation): Load: 500 ohm Pulse Rate: 150Hz Pulse Width: 300 μs
Scope A:
VERT: 20.0V/DIV
HORIZ: 200 μs
OUTPUT: 50.0V pk-
pk
Pulse width: 184 μs
5. S2 MODE(-70% Pulse Width Modulation): Load: 500 ohm Pulse Rate : 150Hz Pulse Width: 300μs
Scope A:
VERT: 20.0V/DIV
HORIZ: 200μs
OUTPUT: 48.8Vpk-pk
Pulse width: 124μs
EMS
1. C MODE : Load: 500 ohm Pulse Rate : 150Hz Pulse Width: 300 μs
Scope A:
VERT:10.0V/DIV
HORIZ: 2mS
OUTPUT:59.1825V
pk-pk
Pulse Rate:148.8Hz
2. S MODE(Synchronous):
Load:500 ohm
Pulse Rate:150Hz Pulse Width:300μs
Contratction Time:12 Sec
Relation Time:12 Sec Ramp Time:6 Sec
MED ITALIA SERVICE S.R.L.S. www.meditaliaservice.com
Scope B:
VERT:10.0V/DIV
HORIZ:50mS
Pulse Rate:5.000Hz
Scope B :
VERT:10.0V/DIV
HORIZ:100μs
OUTPUT:59.1825V
pk-pk
Pulse Width:300μs
Scope B:
VERT:10.0V/DIV
HORIZ:100μs
OUTPUT:49.5000V
pk-pk
Pulse width:150μs
Scope B:
VERT: 20.0V/DIV
HORIZ: 200 μs
OUTPUT: 56.9V pk-
pk
Pulse width: 302 μs
Scope B:
VERT: 20.0V/DIV
HORIZ: 200μs
OUTPUT: 56.9Vpk-pk
Pulse width: 302 μs
Scope B:
VERT:10.0V/DIV
HORIZ:100μs
OUTPUT:59.1825V
pk-pk
PulseWidth:300μs
Manuel de l'Utilisateur
Scope C:
Modulation: -
50%
3. A MODE(Alternate):
Load: 500 ohm Pulse Rate :150Hz
Pulse Width:300μs Contratction
Time :12 Sec Relation Time:12 Sec
Ramp Time: 6 Sec
12