Mains Operating Voltage
Check that the instrument operating voltage marked on the rear panel is suitable for the local
supply. Should it be necessary to change the operating voltage, proceed as follows:
1.
Ensure that the instrument is disconnected from the AC supply.
2.
Remove the screws holding the case upper and handle and lift off the case upper.
3.
Change the transformer connections following the diagram below:
4.
Re-assemble in the reverse order.
5.
To comply with safety standard requirements the operating voltage marked on the rear panel
must be changed to clearly show the new voltage setting.
Fuse
The AC fuse is located in the fuse drawer in the lower part of the IEC inlet connector. To change
the fuse remove the line cord and open the fuse drawer with a suitable tool.
The correct fuse type is 20 x 5mm 250V HBC time-lag with the following rating:
Model
30V/1A
30V/2A; 18V/3.3A; 56V/1.1A
30V/3A; 15V/5A
Dual and Triple
Make sure that only fuses with the required current rating and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuseholders are prohibited.
Mains Lead
When a three core mains lead with bare ends is provided this should be connected as follows:
BROWN
BLUE
GREEN/YELLOW
As the colours of the wires in the mains lead of this apparatus may not correspond with the
coloured markings identifying the terminals in your plug proceed as follows:
The wire which is coloured green-and-yellow must be connected to the terminal in the plug which
is marked by the letter E or by the safety earth symbol shown above or coloured green or green-
and-yellow.
The wire which is coloured blue must be connected to the terminal which is marked with the letter
N or coloured black.
The wire which is coloured brown must be connected to the terminal which is marked with the
letter L or coloured red.
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited.
230V Operation
BLUE
BROWN
-
MAINS LIVE
-
MAINS NEUTRAL
-
EARTH
WARNING! THIS INSTRUMENT MUST BE EARTHED.
115V Operation
BLUE
230V
115V
800mA (T)
1.6A (T)
1.25A (T)
2.5A (T)
1.6A (T)
3.15A (T)
2.5A (T)
4A (T)

Installation

BROWN
Safety Earth Symbol