Capteurs de Pression Piezoresistifs, Pour Applications Moteurs, R&D et Tests, Type 4260/62A...
Tableau 1
Précision de sortie & configuration de sortie électrique
Code
Sortie
A
mV (10 mV/V)
C
4-20 mA
E
4-20 mA, règl.cal ajustable
F
1 ... 6 VDC (3 pôles)
H
1 ... 6 VDC (3 pôles) règl.cal
I
0.5 ... 4.5 VDC (3 pôles)
K
0.5 ... 4.5 VDC (3 pôles) règl.cal
L
0.1 ... 5 VDC (3 pôles)
N
0.1 ... 5 VDC (3 pôles) règl.cal
O
0.5 ... 4.5 VDC (3 pôles) faible puiss, alim. 5 V
Q
0.5 ... 4.5 VDC (3 pôles) faible puiss, alim.
5 V, règl.cal
R
0 ... 10 VDC (4 pôles)
T
0 ... 10 VDC (4 pôles), règl.cal
U
–5 ... 5 VDC (4 pôles)
W
–5 ... 5 VDC (4 pôles), règl.cal
X
0 ... 5 VDC (4 pôles)
Z
0 ... 5 VDC (4 pôles), règl.cal
1
0 ... 5 VDC (3 pôles) avec compensation, règl.cal
2
0 ... 10 VDC (3 pôles) avec compens., règl.cal
3
0.5 ... 4.5 VDC (3 pôles) faible puissance ratio-
métrique
4
–5 ... 5 VDC (3 pôles) avec compens., règl.cal
5
0.5 ... 4.5 VDC (3 pôles) faible puissance ratio-
métrique, règl.cal
6
0 ... 5 VDC (3 pôles) avec compensation, règl.cal
7
0 ... 10 VDC (3 pôles) avec compens, règl.cal
8
–5 ... 5 VDC (3 pôles) avec compensation
Ces informations sont valides au moment de leur publication. Kistler se réserve
le droit de modifier les caractéristiques de ses produits. La responsabilité de
Kistler ne peut être engagée du fait de l'utilisation de ses produits.
SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL: szss20@163.com
Certifications d'Etalonnage
Les certificats d'étalonnage sont disponibles pour tout type
4260/62A et sont fournis en standard pour toute version Pre-
mium.
Les informations suivantes sont alors données sur le certificat
d'étalonnage et permettent un résumé clair des performances
actuelles du capteur par rapport aux spécifications techniques :
• Non Linéarité, Hysteresis, Répétabilité à température ambi-
ante
• Dérive Thermique du zéro et du gain à travers la gamme de
température compensée.
• Résumé des tests auxquels le capteur est soumis
• Détails des raccordements électriques
• Réglage du zéro et du gain à température ambiante
Toute donnée est traçable selon le INST (Institut National des
Standards et de la Technologie) et est certifié ISO17025. Les
types 4260/62A peuvent être fournis avec des étalonnages
spécifiques à la demande selon les applications.
Piezoresistive Pressure Sensor Calibration
Type
Span
4262A150AXCB00W1
V
Serial Number
2050735
Offset
V
Pressure Range
0 to 150 PSI
Supply Voltage
V
Reference
Calibrated By
Gauge
Dan Wolf
Test Condition
Date/Time
New
6/28/2007 8:08:00 AM
Non-Linearity, Hysteresis, and Repeatability (NLHR)
Summary: Passed
Pressure (PSI)
Output (V)
BFSL Error (%)
Limits (%)
Limits (%)
0.0
0.0011
-0.010
± 0.1
Accuracy (NLHR)
± 0.1
37.5
1.2491
-0.002
± 0.1
Temp -10°C To 50°C
± 1.0
75.0
2.4979
0.017
± 0.1
Temp -40°C To 125°C
± 1.5
112.5
3.7455
0.013
± 0.1
Span (5 V nom.)
± 1.0
150.0
4.9916
-0.018
± 0.1
Offset (0 V nom.)
± 1.0
75.0 *
2.4962
-0.034
± 0.1
0.0 *
-0.0007
-0.036
± 0.1
* Decreasing Pressure
Temperature Performance
Temperature (°C)
Span Error (%) Offset Error (%)
-40.2
-0.320
-0.089
-19.1
-0.351
-0.069
-9.3
-0.249
-0.034
-0.5
-0.233
0.005
25.0
0.000
0.000
48.5
0.112
-0.138
77.5
0.131
-0.105
100.5
0.094
-0.018
124.7
-0.098
0.170
Error Calculation
Unit
NLHR limits are based on
% span
Temperature Performance limits are based on
% span
Span & Offset limits are based on
% span
Reference Equipment
Electrical Interface (MIL-C-26482)
Type
S/N
Pin/Wire
Function
Agilent 34970A
MY44021689
A
(+) Supply
Mensor Series 600
610435
B
(-) Supply
D
(+) Output
C
(-) Output
E
N/C
F
N/C
Case
N/C
This Sensor was calibrated per Kistler Test Procedure 680-0000-701 using comparison technique against a Kistler Working Standard. Kistler Working Standards
are periodically calibrated against a Kistler Reference Standard System which in turn is periodically recertified and traceable to the National Institute of
Standards and Technology (NIST). It is derived from accepted values of natural physical constants according to the International System of Units (SI). Kistler's
calibration system meets or exceeds the requirements of MIL-STD-45662A, ANSI/NCSL Z540, ISO 9001:2000 and ISO/IEC 17025. Kistler is accredited to ISO/IEC
17025 by ACLASS, Assured Calibration and Laboratory Accreditation Select Services. Certificate numbers are on file at Kistler and may be requested in writing.
Estimated uncertainity of this calibration is ±0.05 % of pressure range for voltage output sensors or ±0.12% of pressure range for current output sensors with
respect to reference standard. This certificate shall not be reproduced, except in full, without written approval of Kistler Instrument Corporation.
Kistler Instrument Corp.
Tel
1-888-KISTLER
ISO 9001 CERTIFIED QUALITY SYSTEM
040-0090-001
75 John Glenn Drive
Fax
1-716-691-5226
ISO 17025 Accredited Calibration Laboratory
Page 2 of 2
Amherst, NY 14228-2171
info.us@kistler.com
Etalonnage Spécifique
Un étalonnage à la demande permet l'amélioration de la plage
d'erreur statique, des effets thermiques selon la température
d'utilisation et des tolérances sur les paramétrages du zéro et
du gain.
Exemples d'amélioration des erreurs thermiques (dérives ther-
miques du zéro et du gain) :
• 0.5 % FS sur 10 ... 50 °C
• 1.0 % FS sur 10 ... 125 °C
• 1.0 % FS sur –40 ... 50 °C
Merci de contacter Kistler pour toute demande spécifique
d'étalonnage
© 2010, Kistler Instrument Corporation, 75 John Glenn Dr., Amherst, NY 14228
Tel 716-691-5100, Fax 716-691-5226, sales.us@kistler.com, www.kistler.com
4.991
0.001
15.001
Pass/Fail
Passed
Passed
Passed
Passed
Passed