Cleaning process of collection cup and aspi-
ration tip:
Washing the collection cup and aspiration tip
with mild soap in water for 3 minutes. Then
rinse these parts with running hot tap water
for 3 minutes and air dry.
For home use device disinfection, 70% Etha-
nol or Isopropyl alcohol(available in the phar-
macy) can be used. You can clean the col-
lection cup, head module after using several
times. Please prepare a pair of gloves and a
clean cloth or cotton bud that can be moiste-
ned with 70 % alcohol. To clean the entire device, please use a soft cloth
slightly moistened with a mild soapy solution. Under no circumstances
may liquid enter the device. Do not use the device again until it is com-
pletely dry.
NOTE: DO NOT use solvents to clean the unit.
The head module can only be rinsed with water, it may
be rinsed with hot water but must not be cleaned with
any soap or chemical cleaners. Soap and cleaners may
damage the vent which prevents mucus from being
sucked into the device. It is recommended that all pie-
ces be allowed to dry completely before reassembly.
DO NOT wash this device with water or any kind of
liquids. It is strongly recommended to wipe the de-
vice with soft cloth and keep the device clean.
TROUBLESHOOTING
PROBLEM
The suction is
weak or there
is no suction
at all
60
manuale aspiratore_TRAD.indd 60
POSSIBLE CAUSE
Batteries are low or impro-
perly installed.
The grey rubber ring loca-
ted where the collection
cup screws in is missing.
The collection cup is not
properly screwed in.
SOLUTION
Check or replace batteries.
Make sure there is a grey rubber
ring where the collection cup
screws in.
Make sure the three guides on
each side of the collection cup
line up with the guides on the
suction portion of the aspirator.