Télécharger Imprimer la page

Publicité

Liens rapides

HM 110
Quick Start Guide
EN
FR
www.sauermanngroup.com

Publicité

loading

Sommaire des Matières pour sauermann HM 110

  • Page 1 HM 110 Quick Start Guide www.sauermanngroup.com...
  • Page 2 This connection must be made by a qualified and trained technician. To make the connection, the transmitter must not be energized. To make a 3-wire connection, before powering For ACTIVE HM 110-A models with 0-10 V output: up the transmitter, please connect the output ground to the input ground. See drawing below.
  • Page 3 In order to compensate a possible drift of the transmitter, it is possible to add an offset to the displayed value by the HM110 instrument: it shows 48% RH, a standard instrument shows 45% RH. It is then possible, through the software, to integrate an offset of -3 to the displayed value by the HM 110 instrument To access the configuration via software: Modèle arrière...
  • Page 4 Afin de compenser une dérive éventuelle du capteur, il est possible d’ajouter un offset à la valeur affichée ci-dessous. par le HM 110 : il indique 48% HR, un appareil étalon indique 45% HR. Il est alors possible, grâce au logiciel, 4 fils d’intégrer un offset de -3 à...
  • Page 5 Descargue el manual de usuario del software LCC-S Scarica il manuale d'uso del software LCC-S Download the HM 110 data sheet Télécharger la fiche technique du HM 110 Descargue la ficha técnica del HM 110 Scarica la scheda tecnica del HM 110...