meaSUriNg methoDS
in the mouth (oral)
Position the thermometer in one of the two pockets under the tongue, to the left or right of the root of the
tongue. The measuring sensor must be in good contact with the tissue. Close your mouth and breath evenly
through the nose to prevent the measurement from being influenced by inhaled/exhaled air.
Approx. measuring time: 1 minute.
in the anus (rectal)
This is the most reliable measuring method, and is especially appropriate for infants and small children.
Carefully insert the tip of the thermometer 2 to 3 cm into the anal aperture.
Approx. measuring time: 1 minute.
in the armpit (axillary)
From a medical viewpoint, this method will always provide inaccurate readings, and should not be used if
precise measurements are required.
Approx. measuring time: 5 to 10 minutes.
CleaNiNg aND DiSiNFeCtiNg
The thermometer is 100% waterproof and may be immersed in water or disinfectant solution, and wipe off
with a moist towel. Please use one of the following disinfectants for cleaning, following the manufacturer's
instructions in each case.
Name of Disinfectant
Isopropyl alcohol 70%
Note: any other cleaning solution, dilution or cleaning method can cause a malfunction of the
thermometer. the thermometer must never be boiled.
BatterY DiSPoSal
Please break the end cap from the main body with a pair of pliers. Remove the battery and the PC board from
the plastic housing and dispose of separately. Battery cannot be replaced.
teChNiCal SPeCiFiCatioNS
type
measurement range:
measurement accuracy
Self-test
Display
memory
operating environment
Storage temperature
Storage humidity
Battery
Battery lifetime
expected service life
reference to standards
liFetime WarraNtY
This product is warrantied to be free from defects in material or workmanship for the life of the original owner.
It will be repaired or replaced without charge upon receipt of the unit prepaid with $ 5.00 to cover handling,
packaging and return postage to the address below. Please include your full name, address and daytime phone
number or email address.
Thermor Ltd.
Repair Department
16975 Leslie Street
Newmarket, ON L3Y 9A1
Dilution with Water
95.0%
Maximum thermometer
32.0°C to 43.9°C / 89.6°F to 111°F
Temp. < 32.0°C / 89.6°F; display «l» for low (too low)
Temp. > 43.9°C / 111°F; display «h» for low (too high)
±0.1°C / 0.2°F between 34°C / 93.2°F and 42°C / 107.6°F at an
ambient temperature of 18°C / 64.4°F to 28°C / 82.4°F
Automatic internal check at a test value of 37.0°C / 98.6°F; if there is a
deviation of >0.1°C / 0.2°F, «err» (error) is displayed.
Liquid crystal display (LCD) with three digits
Smallest unit of display: 0.1°C / 0.1°F
For storing the last measured value
10°C to 40°C / 50°F to 104°F
-25°C to 60°C / -13°F to 140°F
15% to 95% (noncondensing)
IEC type SR41 Silver oxide high drain
Approx.4500 measurements
2 years
EN12470-3, clinical thermometers
ASTM E 1112, CEI 60601-1, CEI 60601-1-2, CEI 60601-1-11
Type BF applied part
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
Read the instruction manual carefully before using this device,
especially the safety instructions, and keep the instruction
manual for future use.
max. Duration method
Immerse for 24 hours