Chapter 4 Maintenance

Electromagnetic Compatibility

This product has been tested and found to comply with all
applicable requirements of council directive 2014/30/EU and its
amendments. This product has been tested to EN 61326-1:2013
and found to comply with the performance criteria and
requirements of the standard.
This product is intended for use in a Class B electromagnetic
environment such as a biomed laboratory for equipment testing,
maintenance, etc.
This product is not intended for use in the medical / patient
environment. During testing, the product displayed consistent
temperature during testing and did not deviate more than +/-
0.05 deg. C.
If operated under conditions outside the testing parameters:
• If the product experiences a shut down, it is expected that it
can be restored without any errors;
• If the temperature deviates more than +/- 0.05 deg. C during
an EMC event, the product is expected to auto-recover when
the EMC interference stops.
This product can radiate radio frequency energy and if not used in
accordance with the instructions, which may cause harmful
interference to other devices in close proximity. However, there
is no guarantee that interference will not occur in a particular
setting. If this product does cause harmful interference with
other devices, which can be determined by turning the product
on /off, the user is encouraged to try to correct the interference
by one of more if the following measures:
• Reorient or relocate the receiving device
• Increase the separation between equipment