Slide the coils onto the assembly cylinder and
insert them into the basic unit as shown in
Fig. 2.
Take care when doing this to ensure the direc-
tion of the windings of both coils is the same.
The arrows embossed on the coils must point in
the same direction.
Tighten the knurled screws an equal amount by
hand. Make sure that the assembly cylinder is
located accurately on the assembly rings.
Fig. 2 Basic unit complete with coils
3.2 Connection to the control console
Insert the probe into the chamber in the basic
unit so that it touches the housing (see Fig. 3).
Plug the lead from the probe into the socket
"Probe In" on the control console. Take note of
the slot in the connector socket.
Connect the coils to the "Coil" output on the
back of the console.
Connect the control console with plug-in power
supply via socket "12 VAC/1A".
Insert the DPPH sample (orange cap) into the
sample chamber (see Fig. 4).
Fig. 3 Basic unit with probe
Fig. 4 Basic unit with DPPH sample inserted
3.3 Calibration and settings
3.3.1 Use of an oscilloscope
Connect the output "SIGNAL OUT" on the control
console to channel 1 of the oscilloscope and the
output "FIELD OUT" to channel 2 (see Fig. 5).
Set the oscilloscope as follows:
Channel 1: 2 V DC
Channel 2: 1 V DC
Time base: 5 ms
Trigger from channel 2, filter set to Low Fre-
quency
3.3.2 Use 3B NETlog™
Connect the output "SIGNAL OUT" on the control
console to input
and out "FIELD OUT" to input
Connect the 3B NETlog™ unit to a computer and
run the 3BNETlab
Set up a new data record from the "Measure-
ment lab" menu option and define the following
parameters:
Input A: Field, input mode VDC, Input range 2 V
Input B: Signal, input mode VDC, Input range 2 V
Measuring interval: 500 μs (2 kHz)
Click the "Oscilloscope" button and start the
measurement.
An oscilloscope window will open.
Set to trigger from input A, select a falling edge
and a positive trigger threshold at about 10 to
20%.
3.4 Experiment procedure
Set a frequency of about 50 MHz on the control
console. (since the frequency knob is a 10-turn
potentiometer, it may be necessary to turn it
round several times).
Set the sensitivity to where the maximum signal
amplitude is obtained.
2
IN
U
on the 3B NETlog™ unit
B
IN
U .
A
TM
software on the computer.