Filling the Portable System from the Reservoir
1. Check the contents indicator on the reservoir to ensure liquid oxygen is available for
filling purposes. When the reservoir is low, inform your equipment Provider.
2. Remove the cannula from the Device, if attached.
3. Remove the Device from the carrying bag.
4. Remove protective cap on reservoir fill coupler, if applicable.
DO NOT fill the device while it is in the carrying bag.
The Portable System is intended to be used with any compatible reservoir with a
maximum working pressure of 53 PSi (3.65 bar).
5. Using a clean, dry cloth, wipe the fill connector on the reservoir and Device.
6. Carefully position the Portable, ensuring that the fill connector of the Portable System
aligns with the fill connector of the reservoir.
7. Engagement:
PM2200 (Tyco / Puritan Bennett, Top fill)
PM2203 (Taema Topfill):
Connect the portable & reservoir by pressing down to
the fill position, being careful not to depress the
release button on the reservoir.
During filling, maintain a slight downward pressure on
the Portable System with one hand to keep the
device steady and maintain proper filling position.
PM2201 (Mark Series, Top fill with Twist):
Rotate the portable System clockwise until the device
is locked into position (approximate, rotation 90°).
DO NOT USE EXCESS FORCE WHEN LOCKING
INTO POSITION.
PM2202 (Chart / Caire, Side fill):
Rotate the portable System counterclockwise until the
pin of the device engages with the slot of the
reservoir connector (approximate rotation 45°).
Carefully and firmly rotate the portable System back
to the upright position, until the Device and reservoir
are locked together.
continued
WARNING
7