INSULATION RANGES
Nominal test
Voltage (d.c.):
Measuring Range:
Terminal Voltage (d.c.): +15% maximum on open circuit.
Short Circuit Current: < 2 mA
Test Current on Load: 1 mA at min. pass values of insulation specified
Accuracy (at 20 °C):
CONTINUITY RANGES
Measuring Range:
Open Circuit Voltage:
Short Circuit Current: 205 mA, ± 5 mA
Accuracy (at 20 °C):
Zero Offset Adjust:
Continuity Beeper:
RESISTANCE RANGE (can be used for diode testing)
Measuring Range:
Open Circuit Voltage:
Short Circuit Current: 20 µA, ±5 µA
Accuracy (at 20 °C):
TYPICAL TERMINAL VOLTAGE CHARACTERISTICS
VOLTAGE RANGE
Analogue Scale
0 to 500 V
1,0 to 50 V
The instruments meet the requirements for double insulation to IEC 1010-
1 (1995) EN 61010-1 (1995) to Category III*, 300 Volts phase to earth,
440 Volts phase to phase, without the need for separately fused test
leads. If required, fused test leads are available as an optional accessory.
The B M 4 0 2 / 2 and B M 4 0 4 / 2 do not incorporate a voltage range and
must not be intentionally connected to live circuits. *Relates to transient
overvoltage likely to be found in fixed installation wiring.
SPECIFICATION
BM400/2
BM402/2
500 V
1000 V
0,01MΩ to 999 MΩ on all ranges (0 to 10 GΩ
on analogue scale)
in BS7671, HD384 and IEC 364, 2 mA max.
± 2%, ± 2 digits
0,01Ω to99,9 Ω (0 to 50 Ω on analogue scale)
5 V, ±1 V
1Ω to 9,99 Ω: ± 2%, ± 2 digits
10Ω to 99,9 Ω: ± 5%
0Ω to9,99 Ω
Operates at less than 5 Ω
0,1kΩ to100 kΩ (0 to10 MΩ on analogue scale)
5 V, ± 1V
± 5% ± 2 digits
Measurement Scale
0 to 450 V d.c. or a.c. (50/60 Hz)
450 to 600 V d.c. or a.c.(50/60 Hz)
0 to 450 V 400 Hz a.c.
1,0 to 50,0 V d.c. or a.c. (50/60 Hz)
BM401/2
BM404/2
500 V
SAFETY
BM403/2
250 V, 500 V
1000 V
Accuracy(above 1 V)
± 1% ± 2 digits
± 2% ± 2 digits
± 5% ± 2 digits
± 2% ± 3 digits