Appendix
Manufacturer's declaration-electromagnetic emissions
The TH8xyz series is intended for use in the electromagnetic environment specified below.
The customer or the user of the TH8xyz series should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Manufacturer's declaration-electromagnetic immunity
The TH8xyz series is intended for use in the electromagnetic environment specified below.
The customer or the user of the TH8xyz series should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Electrostatic
Contact: 8 kV
discharge(ESD)
Air 2 kV, 4 kV,
IEC 61000-4-2
8 kV, 15 kV
Power frequency
30 A/m
(50, 60 Hz)
50 Hz or 60 Hz
magnetic field
IEC 61000-4-8
Compliance
Group 1
The TH8xyz series uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The TH8xyz series is suitable for use in all establishments,
Class B
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Compliance
level
Contact: 8 kV
Air 2 kV, 4 kV,
8 kV, 15 kV
30 A/m
50 Hz or 60 Hz
Electromagnetic environment – guidance
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
The TH8xyz series power frequency magnetic
fields should be at levels characteristic of a
typical location in a typical home healthcare
environment.