10. Status, Warning and Fault Messages
The following table lists the status, warning and fault messages and their meaning.
10.1 Control Touch Pad:
Element/Message
Explanation
LEDs on the front panel
Fig. 10.1.1
Fault display
The meaning is explained by the corres-
ponding fault message on the LCD
Fig. 10.1.2
Laser output is switched on
Wireless interface connection status (Control Touch Pad to the Laser Unit)
Fig. 10.1.3
No connection to the Laser Unit
Fig. 10.1.4
There is a connection ot the Laser Unit
Fault messages on the LCD
Confirm message with OK
Fig. 10.1.5
Fault in communication
with the Laser Unit
Fig. 10.1.6
Laser power fault in the Laser Unit
Laser power is too low
Fig. 10.1.7
Laser power fault in the Laser Unit
Laser power is too high
Fig. 10.1.8
Temperature at the laser diodes >60°C
The unit switches automatically to a
frequency-modulated operating mode to
protect the laser diodes
Rechargeable battery icons
The battery charging level is displayed
Fig. 10.1.9
Press on the icon to display further infor-
mation about the charging level
Fig. 10.1.10
Batteries are being charged
Fig. 10.1.11
Batteries not inserted or inserted with the
wrong polarity
Action
•
Check the message on the LCD
•
If necessary, switch the unit off and back on
•
Contact the Service team
-
•
Press the button again
•
Switch the device off and unplug it from the mains
•
Then plug the mains power back in and switch
back on
•
Contact the Service team
•
Switch the device off and unplug it from the mains
•
Then plug the mains power back in and switch
back on
•
Contact the Service team
•
Restart laser output
•
If the fault reoccurs, unplug the Laser Unit from
the mains power briefly
•
Then plug the mains power back in and switch
back on
•
Contact the Service team
•
Restart laser output
•
If the fault reoccurs, unplug the Laser Unit from
the mains power briefly
•
Then plug the mains power back in and switch
back on
•
Contact the Service team
•
Insert batteries into the battery compartment
•
Check that the batteries are inserted the right
way round
26
Rechargeable battery status displays: Display when the battery icon is pressed
Fig. 10.1.12
Batteries are not being charged
The battery charging level is displayed as
a percentage value
Fig. 10.1.13
Batteries are being charged
The battery charging level is displayed as
a percentage value
Fig. 10.1.14
Batteries not inserted or inserted with the
wrong polarity
Warning messages when saving a program
Fig. 10.1.15
Saving is not possible as no laser diode
has been selected
Saving not possible as an invalid program
Fig. 10.1.16
name has been entered (either no name
or a name that already exists)
Sound on / off
Fig. 10.1.17
Sound On
Fig. 10.1.18
Sound Off
10.2 Laser Unit:
Element/Message
Explanation
LEDs on the patien component
Fig. 10.2.1
Fault display
The meaning is explained by the corres-
ponding LCD on the Control Touch Pad
Fig. 10.2.2
Laser Unit is switched on
Fig. 10.2.3
Laser output active (switched on)
•
Insert batteries into the battery compartment
•
Check that the batteries are inserted the right
way round
Confirm message with OK
•
Select at least one laser diode
•
Modify the program name
Action
•
Check the message on the LCD
•
Unplug the
Laser Unit from the mains power
•
Then plug the mains power back in and switch
back on
•
Contact the Service team
•
If the green LED does not light up, check whe-
ther the plug-in power supply unit is connected
•
Contact the Service team
27