7. Mount the sensor
Mount the sensor vertically using its support and flanges.
(The TA120 support also allows mounting the sensor horizontally)
TA120 SENSOR ACOUSTIC VERIFICATION
•
Remove the sensor front cover, turn it on and connect it to the PC using the USB-B cable.
•
Run the CESVA sensor manager software, select the port and press "Connect" button.
•
Carefully remove the outdoor kit and place the calibrator correctly.
•
Turn on the calibrator, wait 10 seconds approximately and check that the level is within the range 93.9 ± 0,5 dB.
Otherwise use the + / - buttons until the value is correctly adjusted.
Next press the "Save settings" button.
•
Finally, undo the previous steps.
TA120 SENSOR MAINTENANCE
It is recommended to check the acoustic verification and the TK120 outdoor kit. If the kit is damaged, it must be replaced.
•
Remove the kit carefully.
•
Check the microphone. If it is necessary, remove dust carefully with a brush of very fine camel hair.
•
Carefully to not damage the microphone, insert the outdoor kit.