7. Mount the sensor
Mount the sensor vertically using its support and flanges.
(The TA120 support also allows mounting the sensor horizontally)
TA120 SENSOR ACOUSTIC VERIFICATION
Remove the sensor front cover, turn it on and connect it to the PC using the USB-B cable.
Run the CESVA sensor manager software, select the port and press "Connect" button.
Carefully remove the outdoor kit and place the calibrator correctly.
Turn on the calibrator, wait 10 seconds approximately and check that the level is within the range 93.9 ± 0,5 dB.
Otherwise use the + / - buttons until the value is correctly adjusted.
Next press the "Save settings" button.
Finally, undo the previous steps.
TA120 SENSOR MAINTENANCE
It is recommended to check the acoustic verification and the TK120 outdoor kit. If the kit is damaged, it must be replaced.
Remove the kit carefully.
Check the microphone. If it is necessary, remove dust carefully with a brush of very fine camel hair.
Carefully to not damage the microphone, insert the outdoor kit.