3.2 Symboles
Important safety information. Refer to the manual.
(CATEGORY III MEASURE) fitted to measure mains low voltage connected
CAT III
equipment. (ex : divisional cabinets, electrical cabinets, protections, circuit breakers,
sockets etc...)
Double insulation (protection class II)
Device in which the protection against electric shock does not rely solely on the main
insulation and which includes additional safety measures, such as double insulation or
reinforced insulation, these appliances having no means of protection based on the to
the ground or the conditions of the installation

Direct current

Alternating current
Fuse
Conform to the applicable European directives.
Earth
Hazardous voltage can exists.
Product whose packaging is subject to a sorting order for recycling.
Used electrical products must not be disposed of with household waste. Please use
special facilities to process them. Ask your local authority or sales outlet about
recycling facilities near you.
DISPOSAL OF BATTERIES: This symbol indicates that the batteries and accumulators
supplied with this product should not be treated as simple household waste. To get
rid of them safely, you can either take them to your dealer or drop them at the pickup
points of the point-of-sale stores or your community's garbage collection depots. At
the end of your life, remember to remove the batteries.
Indoor use only.

4. Technical specifications

4.1. General specifications

• Maximum voltage between earth and input : 600V~ou 600V
• Screem : LCD screen 3 1/2 digits
• Calibration : Manual
• Overload protection : protection for all ranges
12
• Hold function
• Relative value measurement
+
• Low battery indicator
• Exceeding the measuring range: «1» appears on the screen.
• Negative Polarity: «-» appears on the screen.
• Operating temperature: 0~40 °C (32~104 °F)
• Storage temperature: -20~60 °C (-4~140 °F)
• Altitude max : 2000m
• Indoor use only
• Power supply: 1.5V×2, AAA R03P battery
• Fuse protection: F 0.5A/500V (fast action) - F 10A/500V (fast action)
• Safety class: EN 61010-1, CAT III 600 V
• Dimensions (L x l x H) : 138 x 67 x 30 mm
• Weight : about 120 g

4.2. Electrical specifications

4.2.1 DC Voltage
Measuring
Resolution Accuracy
range
200 mV
100µV
2 V
1 mV
± 1,0% of measured value + 2 digits
20 V
10 mV
200 V
100 mV
600 V
1 V
± 1,2 % of measured value + 2 digits
4.2.2 Direct current
Measuring
Resolution Accuracy
range
200µA
0,1µA
2mA
1µA
± 1,5 % of measured value + 2 digits
20mA
10µA
200mA
100µA
± 2,0 % of measured value + 2 digits
10A
10mA
± 2,0 % of measured value + 3 digits
4.2.3 AC Voltage
Measuring
Resolution Accuracy
range
200V
100mV
± 1,5 % of measured value + 3 digits
600V
1V
Overload protection:
• 600V
or 600V~
rms for
• Impedance: 1MΩ
other ranges
Overload protection:
• Fuse F 0,5A/500V for
the ranges µA and mA
• Fuse F 10A/500V for
the range 10A
For the range 10A,
measure 10 seconds
max. every 10 minutes.
Overload protection:
600V
or 600V~ rms.
Frequency response:
40 to 500Hz in rms of
a sine wave (average
response).
13