5/6/07
To assemble the column to the base:
1. Identify the parts and lay the base (1) on the floor so that it rests on the three castors.
2. Unscrew and remove the nut and washer from the bolt with sleeve (5).
3. Put the column (2) into the base bracket so the slot and folding handwheel is in the lower front position.
4. Line up the cross hole in the column tube with the hole in the angle bracket and push the bolt and sleeve through both parts.
With the bolt with sleeve in place, replace the washer and nut and tighten them fully with the tools (6) provided.
5. Loosen the folding mechanism handwheel and raise the column into the vertical position. The folding handwheel will drop into
the front slot of the base bracket. Tighten the folding handwheel to secure the column in place.
To assemble the knee pad:
1. Remove the handwheel at the top of the column.
2. Slide the U shaped bracket over the column ensuring the rubber ferrule is at the bottom and the knee pad is facing forwards.
3. Once the knee pad is at the desired height, tighten the knee pad handwheel to secure it in place. Replace the handwheel at the
top of the column.
To assemble the handle bar:
1. Press in the bullet catch that protrudes from the bottom of the handle (4) tube.
2. Slide the handle tube into the main column tube. The bullet catch will engage in the hole in the side of the main column tube.
Please Note: the handle should not be extended beyond this point during use.
3. Press the bullet catch in and lower the handle to the desired height. Holes in the handle tube provide a positive location for the
handwheel. Ensure the handwheel is in a hole and tighten to secure the handle.
To fold the Orbi-Turn:
1. Loosen the folding mechanism handwheel.
2. Raise the handwheel to allow it to clear the base bracket.
3. Lower the column and handle to the floor.
4. To return the handle and column to the upright position, raise the column into the vertical position. The folding handwheel will
drop into the front slot of the base bracket. Tighten the folding handwheel to secure the column in place.

CLEANING AND MAINTENANCE INSTRUCTIONS

Inspect the turner before each use to ensure that it is free from damage and excessive wear. Look particularly for dents, cracks,
loose parts or other damage that may affect the integrity of the product.
It is recommended that surfaces are wiped clean using a wet cloth with a mild detergent or disinfectant. Ensure that the unit is dry
before use.

TECHNICAL SPECIFICATIONS

Constructed from steel for durability and strength.
Unit weight:
Base diameter:
Bottom handle:
Lowest position
Highest position
Top handle:
Lowest position
Highest position
Kneepad (Mid point):
Lowest position
Highest position
Maximum
user weight
Please Note: as the weight of the patient increases the apparent assistance given by the weight and low centre of gravity of the
product will reduce. Consequently the force that the carer must exert will increase.
12:27 pm
Page 4
15kg
400mm
815mm
1175mm
920mm
1315mm
235mm
510mm
190
30
st
kg

Publicité