Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Immunity Test Levels
Professional
Basic EMC
healthcare
standard
facility
environment
Electrostatic
Discharge
±8kV contact
(ESD)
±15kV air
IEC61000-4-2
Electrical fast
transient/
±2kV for power supply line
burst
±1kV for input/output line
IEC61000-4-4
± 1 kV line(s) to
line(s)
Surge
IEC61000-4-5
± 2 kV line(s) to
earth
Voltage dips,
short
Voltage Dips:
interruptions
i) 100% reduction for 0.5 period,
and voltage
ii) 100% reduction for 1 period,
variations on
iii) 30% reduction for 25/30 period,
power supply
Voltage Interruptions:
input lines
100% reduction for 250/300 period
IEC61000-4-1
1
Power
frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80 MHz
Conducted RF
6 Vrms in ISM bands
IEC 61000-4-6
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
HOME HEALTHCARE
ENVIRONMENT
± 1 kV line(s) to
line(s)
30 A/m
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Compliance
Electromagnetic
Levels
Environment-Guidance
Floors should be wood,
concrete or ceramic tile. If
±8kV contact
floors are covered with
±15kV air
synthetic material, the
relative humidity should be at
least 30 %.
±2kV for power
Mains power quality should be
supply line
that of atypical commercial or
±1kV for
hospital environment
input/output line
± 1 kV line(s) to
Mains power quality should be
line(s)
that of atypical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of this device requires
continued operation during
230V
power mains interruptions, it
is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
30 A/m
location in a typical
commercial or hospital
environment.
Portable and mobile RF
communications equipment
should be used no closer to
any part of this device,
6Vrms
including cables, than there
commended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
12