9.2 Changing the arm cuff

If you need a new arm cuff for your "Pressio A", this can be obtained from
Terraillon as an accessory. For technical reasons, the arm cuff should only
be changed by Terraillon. Please consult your dealer if necessary.

10 Descriptions of errors

10.1 Error numbers

Numbers
Causes
Err 0
No pulse or detected pulse is out of range.
Err 1
Arm cuff serious leakage or inflation too low.
Err 2
Systolic cannot be measured.
Err 3
Arm cuff pressure is too high.
Err 4
Diastolic cannot be measured or is outside of range.
Err 5
Deflation is too fast.
Err 6
Deflation is too slow.
Err 7
Systolic outsides the legal range.
Err 8
Mean pressure is outside of range.
Err 9
Arm cuff pressure is not enough even if the highest value is pumped.
28
14:11
Page 28

10.2 Errors classifications and corrections

Conditions
No display after push the
"START/STOP" button
Low battery symbol appears
Battery life is too short.
No reading after measurement
Suspicious measurement results.
Higher or lower than user's
average measurement
Pulse rate higher/lower
than user's average.
Power switches off automatically
Device reinflates during
measurement.
Causes
Corrective Actions
The position of the batteries is
Insert the batteries in the
incorrect
correct position (e.g 5.1)
The voltage of the batteries is
Replace both batteries with
extremely low
new ones.
The voltage of the batteries is
Replace both batteries with
extremely low
new ones.
Warm up the batteries,
When chilly, the batteries might
or place in a warm place for
have a weak electron release
a while then re-start again
May be using wrong battery type.
Use only Alkaline batteries.
The voltage of the batteries is
Replace both batteries with
extremely low
new ones
May not be in correct position
Always take measurements
during measuring
in correct position (e.g 6)
Relax and try to
The blood pressure naturally
measure again
varies from time to time
Move arm or body during
Do not move arm or body
measurement
during measurement
Do not measure right
Measuring right after exercice
after exercice
Push the
"START/STOP"button
System design
again, and then
measure again
This is normal: if user's blood
pressure exceeds the initial
pressure value, the device will
Relax and try to
automatically pump to a higher
measure again
pressure, by 20mmHg each time,
until it reaches a suitable level.
The arm cuff is not fastened
Fasten the arm cuff again
29