Electromagnetic Compatibility (Emc) Tables - DJO cefar rehab X2 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

14. ELECTROMAGNETIC COMPATIBILITY
(EMC) TABLES
34
The CEFAR REHAB X2 Electrotherapy System is intended for use in the electromagnetic environment deified below. The customer
or the user of the CEFAR REHAB X2 Electrotherapy System should assure that it is used in such an environment
Emissions Tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
IEC 61000-3-3
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Compliance
Group 1
Class B
Not Applicable - Battery powered
Not Applicable - Battery powered
Electromagnetic Environment - Guidance
The CEFAR REHAB X2 Electrotherapy System uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby
electronic equipment.
CEFAR REHAB X2 Electrotherapy System is suitable for use
in all establishments, including domestic establishments and
those directly connected to the public low voltage power
supply network that supplies buildings used for domestic
purposes

Publicité

Table des Matières
loading

Produits Connexes pour DJO cefar rehab X2

Table des Matières