Appendix 1 Guidance and Manufacturer Declaration Tables
Guidance and manufacturer's declaration–electromagnetic
emissions
The Model
THERMO FH2 Infrared Thermometer
the electromagnetic environment specified below. The customer or the
user of the Model THERMO FH2 Infrared Thermometer should assure
that it is used in such an environment.
Compliance
Emissions
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic
N.A.
emissions
IEC 61000-3-2
Voltage
N.A.
fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic
The
Model THERMO FH2 Infrared Thermometer
the electromagnetic environment specified below. The customer or the
user of the Model THERMO FH2 Infrared Thermometer should assure
that it is used in such an environment.
Immunity
IEC 60601
test level
test
±8 kV contact
Electrostatic
±2 kV, ±4 kV,
discharge
±8 kV, ±15KV
(ESD)IEC
air
61000-4-2
Power
30 A/m,
frequency
50/60Hz
(50/60 Hz)
magnetic
field IEC
61000-4-8
NOTE U is the a.c. mains voltage prior to application of the test level
T
13
is intended for use in
Electromagnetic
environment-guidance
The
Model THERMO FH2 Infrared
Thermometer uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
The
Model THERMO FH2 Infrared
Thermometer is used in home
and it's powered by DC 3V
immunity
is intended for use in
Electromagnetic
Compliance
environment-guidance
level
±8 kV contact
Floors should be wood,
±2 kV, ±4 kV,
concrete or ceramic tile.
±8 kV, ±15 KV
If floors are covered with
air
synthetic material, the
relative humidity should
be at least 30 %.
Power frequency
30 A/m,
magnetic fields should
50/60Hz
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
Guidance and manufacturer's declaration – electromagnetic
immunity
The
Model THERMO FH2 Infrared Thermometer
the electromagnetic environment specified below. The customer or the
user of the Model THERMO FH2 Infrared Thermometer should assure
that it is used in such an environment.
Immunity
IEC 60601
Compliance
test
test level
level
3 Vrms150
Conducted
N/A
kHz to 80
RF IEC
MHz 6 Vrms
61000-4-6
150 kHz to
80 MHz
outside ISM
bandsa
Radiated
10 V/m
RF IEC
80 MHZ to
10 V/m
61000-4-3
2.7 GHz
is intended for use in
Electromagnetic environment
- guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Model THERMO FH2
Infrared Thermometer, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
é
ù
3
5 .
d
=
P
ê
ú
V
ë
û
1
é
ù
3
5 .
d
=
P
80MHz to 800MHz
ê
ú
E
ë
û
1
é
ù
7
d
=
P
800MHz to 2.7GHz
ê
ú
E
ë
û
1
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres(m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
a
survey, should be less than
the compliance level in each
b
frequency range
Interference may occur in the
vicinity of equipment marked
with the following symbol:
14