1. Locate the knurled/thumb screw on the left side of the
header panel.
2. Remove the knurled/thumb screw by hand.
3. Once the screws are removed, wear hand gloves and
gently pull the bracket out until you see the batteries
completely.
4. Replace the batteries and gently close the bracket. Once
the bracket is closed, Align the holes on the bracket and
header panel, screw using the knurled/thumb screw.
Preparation for Storage
If the unit is going to be stored in an off condition, allow the
unit to warm up and dry out with the door open before moving
into storage.
24 |
Maintenance
Cleaning the Unit
(Chromatography
Refrigerators)
Use a solution of water and a mild detergent for cleaning.
Lightly spray the interior storage components and wipe them
dry with a soft cloth or spray the cloth first and then wipe
interior surfaces.
Do not spray or pour directly into holes and gaps in the
chromatography safety circuit assembly for cleaning. Use a
wet wipe to clean around sensor box and outlet.
If the sensor led turns red after cleaning, manually reset the
safety circuit assembly using the manual reset switch on the
right side of header panel.
Replacing Sensor
(Chromatography
Refrigerators)
The sensor needs replacement if the temperature goes below
freezing in the cabinet. Use system alarms to ensure the
temperature inside the unit is always above 0°C. If exposed to
freezing conditions the sensor should be replaced.
If the display shows an error and sounds an alarm (in case of
a control, defrost, bottle or ambient probe failure), the sensor
needs to be replaced.
Sensor Maintenance
(Chromatography
Refrigerators)
The sensor should be replaced every five years by a trained
service provider.