T C - 9 D O P E R A T O R ' S M A N U A L
OPERATION
TC-9D
Use the up and down arrows to enter a
1
The bath will begin to heat immediately. The top display is the
measured thermocouple temperature and bottom is the setpoint.
To change the display from
2
the 2nd button from the left until the lower display shows "UNITS".
Next press Up arrow to reach preferred setting. Press 1st putton
on the left to exit after changes are made.
Two setpoint programming
3
at menu option "SP1" and "SP2". See step 2 to access options
and exit out. Menu option "SP.SEL" tells the controller which to use.
calibration offset
A single point
4
option "PV.OFS". See step 2 to access options and exit out.
Set a correction value in the opposite direction of your external
reference reading. i.e. - reference shows the bath to be 3.0
degrees high, then set "PV.OFS" to "-3.0"
setpoint temperature
oC to oF
and vice versa press
can be used by entering values
can be applied by accessing
15
.