Télécharger Imprimer la page

Diagnostic Management - Aglient TwisTorr 304 FS AG Mode D'emploi

Publicité

17
Technical Information
Profibus Option
7
2
mA
9
2
W
11
2
W
13
2
Hz
15
2
Hz
17
2
-
19
2
min
21
2
h
23
4
mBar/Torr/
Pa

Diagnostic Management

In addition to the ProfiBus six byte standard diagnostic, the
interface provides both user diagnostic and static diagnostics
functions
The standard diagnostic management is fully compliant with the
ProfiBus specification.
During the start-up phase after power on, the first byte in the
user diagnostic area, can assume values different from zero.
This byte has the following meaning:
288/301
0-10000
Current
Motor current consumption
0-200
Power
Motor power adsorption
0-200
Power Limit Applied
Maximum allowable power
0-1500
Driving Frequency
Read the rotational speed of the pump
0-1500
Target Frequency
Read the target rotational speed
0-65536
Cycle Number
Number of cycles (start and stop) done by the pump
0-65536
Last Cycle Time
How long the last cycle lasted
0-65536
Pump Life
Total pump running time
-
Pressure Reading
Reports the pressure reading in the "pressure unit"
selected by Par. Buff., Offset-20, B
Bit 0 = 1: controller not responds (RS485 link broken)
Bit 1 = 1: controller not identified (not compatible with the
ProfiBus gateway)
TwisTorr 304 FS AG Rack Controller User Manual / 87-901-029-01
200
202
155
203
120
301
300
302
224
5-6

Publicité

loading

Produits Connexes pour Aglient TwisTorr 304 FS AG

Ce manuel est également adapté pour:

X3506-64002X3506-64003