Acid Demand To Ph 4.3 With Tablet - Aqualytic AL200 Mode D'emploi

Masquer les pouces Voir aussi pour AL200:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 52
Methods
GB
S:4.3
0.0.0
Zero
Test
S:4.3
RESULT
Notes:
1. The terms total Alkalinity, Alkalinity-m, m-Value and Alkalinity to pH 4.3 are identical.
2. For accurate results exactly 10 ml of water sample must be taken for the test.
3. Conversion table:
Acid demand
to pH 4.3
in mmol/l
x Factor = Result
* Carbonate hardness (reference = Hydrogencarbonate-anions)
Example: 2.5 mmol/l · 2.8 = 7.0° german Hardness (°dH)
Reagent
ALKA-M-PHOTOMETER
42

Acid demand to pH 4.3 with Tablet

0.1 – 4 mmol/l
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Add one ALKA-M-PHOTOMETER tablet straight from the foil to
the 10 ml water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display as Acid demand to pH 4.3 in
mmol/l.
Tolerance: ± 5% (full scale)
CaCO
3
Calciumcarbonate
german Hardness
Factor
50
Form of reagent/Quantity
Tablet / 100
°dH*
°fH*
french Hardness
Factor
Factor
2.8
5.0
marks
°eH*
english Hardness
Factor
3.5
Order-No.
4513210BT
AL200_8 12/2016