Maintenance And Cleaning; Monitoring And Maintenance - Ivoclar Vivadent VP3 easy Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 23
7.

Maintenance and Cleaning

For safety reasons, disconnect power plug before
maintenance and cleaning.
7.1

Monitoring and Maintenance

The time for maintenance procedures depends on the
frequency of use and the location of the pump. For that
reason the recommended times are only approximates.
What
Part
Check all plug-in
Connections
connections for
correct fit.
Check the hose
Vacuum hose
to make sure that
no condensation
water has
accumulated in
the system.
This apparatus has been developed for typical use
in the dental laboratory.
If the product is used in a production facility, for industrial
applications, or in continuous operation, premature ageing
of certain spare parts have to be expected.
These spare parts are e.g.:
Sound absorber
Diaphragm
Valve reed
These spare parts are not covered by the warranty.
Please also observe the shorter service and maintenance
intervals.
10
7.2
Remove dust from the pump from time to time. Use a
vacuum cleaner with a cleaning brush for that purpose.
When
Weekly
7.3
Weekly
Please observe the relevant rules and regulations for the
prevention of accidents as well as the safety and medical
regulations. The connection cover that is marked with the
label shown below may only be opened by qualified per-
sonnel after the power has been disconnected.
Cleaning
Item
Frequency
Housing
if required
Sound absorber
if it is dirty
(4)
Notes on maintenance
The diaphragm may only be changed by an
Ivoclar Vivadent Service Center. The fuse must be
changed by qualified personnel.
Cleaning material
soft, dry cloth
sound absorber
has to be
replaced

Publicité

Table des Matières
loading

Table des Matières