The PiKo stores the last 96 test values and other data associated with
the tests in its memory. If the memory becomes full with 96 tests and
another test is done, the oldest test will be eliminated from memory.
The newest test will then be number 96.
To scroll through the memory:
1. Press and hold the operate button for 5 seconds. The display will
show the last test results containing the PEF (PiKo-1 only), FEV1, FEV6
(PiKo-6 only), FEV1/FEV6 ratio (PiKo-6 only), colour zone and the Q
factor, if any. It will also show the respective memory counter location.
2. Release the button. Each additional press of the button will now
display the next oldest stored test.
To exit the Memory Review Mode either press and hold the operate
button for 5 seconds or avoid pressing the operate button for 20
seconds. The PiKo will then display the latest test result and will be
ready for a new test.
All the test data stored in the PiKo memory can be downloaded to a PC
using the optional PiKoNET Software (interface cradle included).
This software can be purchased on the www.nspirehealth.com website.
NOTE: For the PiKo-1 the entire memory can be erased by setting the
PEF reference to 010, or the FEV1 reference to 0.10. For the PiKo-6, the
entire memory can be erased by setting the FEV1 or the FEV6 reference
value to 0.10 or the FEV1/FEV6 ratio to 0.01.