Hach HART Manuel D'utilisation page 11

Module
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 12
Option
Description
Module
Displays information about the installed module.
information
• Software version
• Bootloader version
• Serial Number
Default Setup
Sets the configuration to factory defaults.
Troubleshooting
For general problems with HART module communications, try the
corrective actions listed below. If the problem continues or other
problems occur, contact technical support.
• Make sure the connections between the module and the HART host
are correct.
• Make sure the loop supply is live.
• Make sure the HART host is set to the correct device address.
• Check the module network setup for the selected channel. Make sure
that the installed device is mapped to the selected channel. Make sure
the SET LOW VALUE and the SET HIGH VALUE are within the range
of the selected source.
• If the HART output from the HART host cannot be trimmed or fixed,
make sure the sensor does not have an error that makes the
controller hold or transfer the output at a fixed value.
English 11