Operation
Using skin temperature control
The system initially proposes the default temperature
setting 36.3 °C.
1 The value 36.3 °C flashes in the display.
2 Briefly press and release p or P to confirm the
value.
1 The displayed value remains constant.
The temperature setting can be adjusted in 0.1 °C steps:
2 Hold down key p or P until the desired temperature
setting appears in the display.
O
3 The yellow LED
lights up or flashes according to
the adjustment (heater function display).
To check heater function during operation:
4 Press the 6 key.
5 The simulated temperature value must be 36 °C
(±0.1 °C) (function check).
If the sensor probe is disconnected or the sensor is
defective:
6 Three flashing dashes in the middle of the display
7 the red alarm LED marked Sensor flashes.
A continuous audible alarm tone is emitted.
Immediately connect the sensor plug or change the
skin temperature sensor.
If the error cannot be remedied immediately:
Remove the skin temperature sensor from the patient.
Switch to manual mode (»Man«).
See "Troubleshooting", page 50.
42
2
2
1
5
4
6
7
3