2. Set the switch on the hFE setting.
3. Find out whether this transistor is a NPN or PNP
transistor and connect the Emitter- (e), Basic (b)
and collector- (c) cable to the appropriate socket in
the switch gear panel on the front side.
4. The display shows the (approximate) hFE value for
a basic current of 10 ηA, V CE 2.8 V.
10.7 MEASURING ALTERNATING CURRENT
(AAC)
1. Connect the BLACK lead to the COM connector
socket. Connect the RED lead to the «VΩmA» con-
nector socket. For measurements between 200 mA
and 20 A, remove RED lead to «20 A» connector
socket.
2. Set the FUNCTION switch on the desired «A~» set-
ting and connect the leads in series to the power
circuit to be measured. The polarity of the measur-
ing tips is not relevant.
3. Read off the value in (milli-) Ampere.
RANGE
2 mA
20 mA
200 mA
20 A
Overload protection 0.2 AI250-V cut-out, 20-A-range
not protected. Frequency range: 40-1000 Hz. Display:
Corrected centre point (corresponds to the effective
value in a sinusoidal form.)
HINT:
1. In case the current range is previously unknown,
you must set the FUNCTION switch in the highest
measurement range and then reduce the measure-
ment range gradually if required.
2. If the fi gure «I» now appears on the display, the
measurand lies outside the range set. The FUNC-
TION switch must be switched to a higher meas-
urement range.
3. The 20 A range is not protected with a fuse. Do not,
therefore, measure for longer than 10 seconds.
10.8 CONTINUITY TEST
1. Connect the BLACK lead to the «COM» connector
socket and the RED lead to the «VΩmA» connector
socket. (Hint: The polarity of the red lead is then
«+».)
2. Set the FUNCTION switch on the «
connectthe measuring tips to the power circuit to be
measured.
3. If the resistance in the power circuit is lower than
50 Ω, the buzzer sounds.
11. SERVICING
CAUTION
•
Always remove the battery before servicing and
cleaning of the multimeter. Never use water or
other fl uids to clean the device.
•
Keep the instrument leads and the multimeter
clean. Some cleaning media and solvents (pet-
•
•
12. PROBLEMS
In case the multimeter does not operate properly, the
cause could be one of these:
1. The multimeter does not show anything.
• The battery is fl at.
- Change the battery.
• The current or the voltage measured was too high
- Contact the Service address in case of problems.
ACCURACY
2. The multimeter does not show any measured
values.
• The fuse has burnt.
- Replace the fuse.
± 2,5%
• One or more leads are defective.
- Replace the instrument leads.
• The current or the voltage measured was too high
- Contact the Service address in case ofproblems.
REPLACING THE FUSE
1. Set the switch on the OFF setting.
2. Remove the rear panel of the multimeter by loosen-
CAUTION! The device must be switched off and the
leads removed before opening the casing of the mul-
timeter.
3. Remove the defective fuse and replace it with a fuse
13. CHANGING BATTERY
1. A battery symbol appears on the display if the bat-
2. Set the rotary knob on the OFF setting.
3. Remove the rear panel of the multimeter by loosen-
» range and
CAUTION! The device must be switched off and the
leads removed before opening the casing of the mul-
timeter.
4. Replace the 3x 1,5 Volt battery by 3xAAA 1,5 Volt.
rol, thinner etc.) can attach or dissolve the plastic.
These products contain benzol, trichloroethane,
chlorine, aqueous ammonia etc.
Clean the casing regularly using a soft piece of
cloth, preferably after every use.
Remove the stubborn dirt using moist cloth. Do not
use any solvents such as petrol, alcohol, ammoniac
solution etc. Such substances damage the plastic
parts.
and the multimeter was damaged in spite of the
safety devices.
and the multimeter was damaged in spite of the
safety devices.
ing the screws.
of the same amperage and tripping characteristics
(250V~F200mAL).
tery is almost discharged.
ing the screws.
GB
11