Standard Calibration; Weight Calibration; Zero Calibration; Span Calibration - Gram SC2 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ZP 5: 5 division not to display at zero
3. Press the

STANDARD CALIBRATION

1. Press and hold the
2. Use the
and
ECF-1 – Weight calibration
1. Press the
key to calibration the zero weight after stable.
2. Use the 0 – 9 keys to enter the weight to be calibrated.
3. Place the required calibration weight on the platform.
4. Press the
5. The calibration procedure is completed.
6. Press the
ECF-2 – Zero calibration
1. Press the
Key to calibration the zero weight after stable.
2. Press the
ECF-3 – SPAN calibration
1. Press the
2. Use 0 – 9 keys to enter the weight to be calibrated.
3. Place the required calibration weight on the platform.
4. Press the
5. The calibration procedure is completed.
6. Press the
to confirm
key, then press the
keys to select ECF-1, ECF-2 or ECF-3
to enter, display will show the internal value, press again the
key to confirm.
key to exit and return to previous setup mode.
key to enter, display will show the internal value, press again the
key to exit and return to previous setup mode.
key to enter.
key to confirm.
key to exit and return to previous setup mode.
key.
11

Publicité

Table des Matières
loading

Table des Matières