  • Page 1 MICROPIPETAS AUTOCLAVABLES AUTOCLAVABLE MICROPIPETTES MICROPIPETTES AUTOCLAVABLES DIGIPETTE MANUAL DE INSTRUCCIONES ..02 INSTRUCTION MANUAL ... . .08 MANUEL D’UTILISATION ... .14...
  • Page 2 Todas las pipetas disponen de eyector de pun- tas integrado. Cubren un rango de volumen de 0.1ul-10000ul. pipetas de volumen variable Referencia Volumen 66732020 0,1-2.5 µL 66732110 0.5-10 µL 66732120 2-20 µL 66732150 5-50 µL 66732210 10-100 µL...
  • Page 3 TÉCNICA DE PIPETEADO A. Pipeteado directo 1. Pulse el botón hasta el primer tope. 2. Sumerja la punta colocada en la boquilla de la pipeta en la solución hasta una profundidad de unos 2-3 mm, y suelte suavemente el botón. Es- pere unos segundos y retire la pipeta tocando el borde del tubo para eliminar el exceso de líquido adherido a la superficie externa de la punta.
  • Page 4: Recomendaciones Para El Pipeteado

    RECOMENDACIONES PARA EL PIPETEADO Aspire líquido únicamente cuando una punta ha sido colocada en la boquilla de la pipeta. Permita que la solución, puntas y pipetas se equilibren con la temperatura ambiente. Se recomienda enjuagar la punta unas 5 veces con el líquido que va a ser dispensado.
  • Page 5 Mantenimieno interno 1. Mantenga pulsado el eyector de puntas. 2. Coloque el diente de la herramienta de apertura entre el eyector de puntas y el cuello del eyector de puntas para li- berar el mecanismo de blo- queo. 3. Suavemente libere el eyector de puntas y retire el cuello del eyector.
  • Page 6: Almacenamiento

    Notas importantes l. El procedimiento se debe llevar a cabo a una temperatura constante de 20º (+/- 0.5º C). 2. El vaso de pesada, agua destilada, pipeta y puntas deberán estar a la misma tempera- tura. 3. Utilice una balanza analítica con una re- solución de 0.01 mg.
  • Page 7 —Punta no adecuada. Utilice otra punta. —Presencia de partículas entre la punta y la bo- quilla. Limpie la boquilla. —Insuficiente cantidad de grasa entre el piston y la arandela. Limpie y engrase la arandela. —Arandela en posición incorrecta o dañada. Cambie la arandela.
  • Page 8 All pipe es are equipped with a built-in tip ejector. They cover the volume range at 0.1ul-10000ul. variable volume pipettes Reference Volume 66732020 0,1-2.5 µL 66732110 0.5-10 µL 66732120 2-20 µL 66732150 5-50 µL 66732210 10-100 µL...
  • Page 9: Pipetting Techniques

    PIPETTING TECHNIQUES A. Forward Pipe ing 1. Press the operating bu on to the first stop. 2. Dip the tip a ached to the pipe e into the so- lution to a depth of about 2-3 mm, and slowly re- lease the operating bu on.
  • Page 10: Pipetting Recommendations

    PIPETTING RECOMMENDATIONS Aspirate liquid into the pipette only when a tip is attached to its tip cone. Allow liquids, tips, and pipettes to equilibrate to the ambient temperature. Pre-rinsing of tip 5 times with the liquid to be dispensed is recommended. This is important es- pecially when dispensing liquids which have a viscosity and density different from water.
  • Page 11 In-House maintenance 1. Hold down the tip ejector. 2. Place the tooth of the ope- ning tool between the tip ejec- tor and the tip ejector collar to release the locking mechanism. 3. Carefully release the tip ejector and remove the ejector collar.
  • Page 12 Important Notes l. Procedure should take place at 20° (+/- 0.5 C) constant temperature. 2. The weighing beaker, distilled water, pi- pette & tips must be at the same temperature. 3. Use an analytical balance with 0.01 mgs readability. 4. Pre-rinse the tip 3 to 5 times before pipet- ting.
  • Page 13 —Insufficient amount of grease on piston and O-ring: clean and grease O-ring. —O-Ring not correctly positioned or dama- ged: change the O-ring. 3. Inaccuracies: —Incorrect operation: follow instructions ca- refully. —Calibration altered: recalibrate according to instructions. —Unsuitable for the particular liquid pipetting technique: use correct pipetting technique.
  • Page 14 Toutes les micropipe es disposent d’un éjecteur de pointes intégré. Elles couvrent une gamme de volume de 0.1 µl à 10000 µl. micropipettes à volume variable Référence Volume 66732020 0,1-2.5 µL 66732110 0.5-10 µL 66732120 2-20 µL 66732150 5-50 µL...
  • Page 15 TECHNIQUE DE PIPETAGE A. Pipetage directe 1. Appuyez sur le bouton afin d’arriver jusqu’au premier cran. 2. Plongez le bout de la pointe de la micropipe e dans la solution jusqu’à une profondeur d’envi- ron 2-3 mm, er relâchez doucement le bouton. At- tendez quelques secondes et retirez la micropipe e en touchant le bord du contenant pour éliminer l’excès de liquide adhérant à...
  • Page 16: Recommandations Pour Le Pipetage

    RECOMMANDATIONS POUR LE PIPETAGE Aspirez le liquide uniquement lorsque la pointe est positionnée sur l’embout de la micro- pipette Faites en sorte que la solution, les pointes et la micropipette s’équilibrent avec la température ambiante Il est recommandé de pré-rincer 5 fois la pointe avec le liquide qui va être distribué.
  • Page 17: Il N'est Pas Necessaire De Recali- Brer La Micropipette Apres Auto- Clavage

    Entretien interne 1.Maintenez appuyé l’éjec- teur de pointes. 2. Placez la dent de l'outil permettant l’ouverture entre l'éjecteur de pointes et le col d'éjection pour libérer le méca- nisme de verrouillage. 3. Libérez en douceur l’éjec- teur de pointes et retirer le col de l’éjecteur.
  • Page 18: Resolution De Problemes

    2. Le bécher de pesée, l’eau distillée, la mi- cropipette et les pointes doivent être à la même température. 3. Utilisez une balance analytique de préci- sion avec une résolution de 0.01 mg. 4. Rincez la pointe de 3 à 5 fois avant de pro- céder au pipetage.
  • Page 19: Garantie De 3 Ans

    —Quantité insuffisante de graisse entre le piston et la rondelle: nettoyez et graisser la rondelle: ron- delle en mauvaise position ou endommagée: remplacez la rondelle. 3. Imprécisions: —Mauvaise utilisation de la micropipette: suivez précisément le manuel d’utilisation. —Calibration altérée: recalibrez la micropipette selon le manuel.